Good Enough for Police v. Good Enough for Charter section 7, 8, and 11(d)


Dry Gas cannister attached to Intoxilyzer
Dry Gas (Air Gas) Control Test System on Intoxilyzer 8000 challenges the Intoxilyzer and Technician with a yes or no decision


I agree that single point control tests and the Internal Test Procedure (ITP) check are good practice for police. The instrument shutting down, if the instrument detects an anomoly during a control test, or an ITP, is good design. However, these safeguards alone are not enough to ensure a scientifically reliable measurement result FOR A FORENSIC PURPOSE.

It is imperative that lawyers learn the differences between calibration of a measuring device at the factory or authorized service centre and control tests at time of use. The latter are very helpful in screening out bad tests, only if the device has been properly maintained, and the control test is done properly, using a reliable alcohol standard.



wet-bath simulator attached to 8000C
Wet-bath Simulator Control Test System on Intoxilyzer 8000C challenges the Intoxilyzer and Technician with a yes or no decision

The problem is that although a single point control test, out in the field, should be good enough for a police officer yes or no decision, such limited evidence of reliability is not adequate for the forensic assessment of scientific reliability of a quantitative analysis. Scientific reliability of a quantitative analysis measurement result is a concept central to both the Motherisk Inquiry Report and the decision of the SCC in St-Onge Lamoureux. Scientific reliability of a quantitative analysis is established through, recent calibration as in Jackson, or recent re-calibration as in Vallentgoed, combined with strict adherence to standard operating procedures i.e. good laboratory practice.


Our job as defence counsel is to convince Canadian criminal courts, that Charter sections 7, 8, and 11(d) require that the Criminal Code of Canada, and other criminal statutes, that define delicts in terms of quantities and units, be construed and applied in accordance with good measurement science.

Comments


If you are a member of the public, please don't attempt to use what you see or read at this site in Court. It is not evidence. The author is not a scientist. The author has a great deal of experience in cross-examining scientists about these issues, but the author is not a scientist. Hire a criminal lawyer in private practice in Ontario. Your lawyer can retain an expert. The author is a retired lawyer, not a lawyer in private practice. Read the statement of the purpose of this web site below.

© 2025 Allbiss Lawdata Ltd.

This site has been built by Allbiss Lawdata Ltd. All rights reserved. This is not a government web site.

For more information respecting this database or to report misuse contact: Allbiss Lawdata Ltd., Mississauga, Ontario, Canada, 905-273-3322. The author and the participants make no representation or warranty  whatsoever as to the authenticity and reliability of the information contained herein.  WARNING: All information contained herein is provided  for the purpose of discussion and peer review only and should not be construed as formal legal advice. The authors disclaim any and all liability resulting from reliance upon such information. You are strongly encouraged to seek professional legal advice before relying upon any of the information contained herein. Legal advice should be sought directly from a properly retained lawyer or attorney. 

WARNING: Please do not attempt to use any text, image, or video that you see on this site in Court. These comments, images, and videos are NOT EVIDENCE. The Courts will need to hear evidence from a properly qualified expert. The author is not a scientist. The author is not an expert. These pages exist to promote discussion among defence lawyers.

Intoxilyzer®  is a registered trademark of CMI, Inc. The Intoxilyzer® 5000C is an "approved instrument" in Canada.

Breathalyzer® is a registered trademark of Draeger Safety, Inc., Breathalyzer Division. The owner of the trademark is Robert F. Borkenstein and Draeger Safety, Inc. has leased the exclusive rights of use from him. The Breathalyzer® 900 and Breathalyzer® 900A were "approved instruments" in Canada.

Alcotest® is a registered trademark of Draeger Safety, Inc. The Alcotest® 7410 GLC and 6810 are each an "approved screening device" in Canada.

Datamaster®  is a registered trademark of National Patent Analytical Systems, Inc.  The BAC Datamaster® C  is an "approved instrument" in Canada.

bottom of page